Equipment:
- Simultaneous display of pH or mV value and temperature
- Stability indicator prevents faulty measurements
- Automatic 1/2-point calibration with 5 pre-programmed standard buffers (pH 4.01; 6.86; 7.01; 9.18; 10.01)
- Automatic or manual temperature compensation
- Graphic symbols show the user how to calibrate and use the unit
Additional features of HI 2215-02
- Integrated measured data memory
- USB port
- Automatic 1/2/3/4/5 point calibration with 7 pre-programmed buffers (pH 1.68; 4.01; 6.86; 7.01; 9.18; 10.01; 12.45) and 2 user-specific buffers
Technical specifications:
| Type |
HI 2211-02 |
HI 2215-02 |
| Measurement range |
pH mV
°C |
-2.00 to 16.00 ±399.9 (ISE) ±2000 (ORP) -9.9 to +120.0 |
-2.00 to 16.00; -2.000 to 16.000 ±399.9 (ISE) ±2000 (ORP) -9.9 to +120.0 |
| Resolution |
pH mV
°C |
0.01 0.1 (ISE) 1 (ORP) 0.1 |
0.01; 0.001 0.1 (ISE) 1 (ORP) 0.1 |
| Accuracy |
pH mV
°C |
±0.01 ±0.2 (ISE) ±1 (ORP) ±0.5 |
±0.01; ±0.001 ±0.2 (ISE) ±1 (ORP) ±0.5 |
| Measurement data memory |
- |
100 measured values |
| Dimensions (L x W x H) |
240 x 182 x 74 mm |
| Weight |
2.5 kg |